Pacific Xtreme PTA Balloon Catheter  

Percutaneous Transluminal Angioplasty

Overview

Treat longer lesions with the Pacific™ Xtreme PTA balloon catheter OTW 0.018" that is offered also in lengths from 150 to 300 mm. These balloons are well suited for use in treating long lesions in the femoral and popliteal vessels.

Indications

The Pacific Xtreme is indicated for percutaneous transluminal angioplasty (PTA) in patients with obstructive disease of peripheral arteries, i.e. ilio-femoral-, femoral-, popliteal-, infra-popliteal- and renal arteries.

Pacific Xtreme for carotid and supra-aortic applications

Only balloon catheters with diameters from 2 mm to 7 mm and lengths from 20 mm to 40 mm are indicated for carotid and supra-aortic applications, with the exception of catheters with a usable length of 180 cm.

Contraindications

The Pacific Xtreme PTA catheter is contraindicated for use in coronary arteries.  The Pacific Xtreme PTA catheters with diameters from 2 mm to 7 mm and balloon lengths of more than 40 mm with usable lengths of 90 cm and 130 cm are contraindicated for use in carotid and supra-aortic arteries.  The Pacific Xtreme PTA catheter with a usable length of 180 cm is contraindicated for use in carotid and supra-aortic arteries.

Product Details

Proprietary Shaft Design

  • For strong pushability and the kink resistance, combined with excellent flexibility

Low Profile

  • 4 F introducer sheath compatibility for less possible puncture trauma
  • Low crossing profile, reliable rewrap and safe retrieval even after multiple inflations*

Dedicated Flexitec Xtreme Balloon Material

  • For high flexibility while maintaining high RBPs

Extreme RBPs

  • For highly calcified lesions

Proprietary Technology

  • To reduce the "banana effect"

3 Shaft Lengths (90 cm, 130 cm and 180 cm)

  • For more access site options

Manuals and Technical Guides

Instructions for Use

Find this technical manual in the product labeling supplied with each device.

Model Specifications

Type Specification
Catheter Design Over the wire (OTW), coaxial shaft
Balloon Material Flexitec™ Xtreme
Balloon Coating LFC hydrophilic
Balloon Marker 2 swaged, Platinum Iridium
Usable Shaft Lengths 90, 130, 180 cm
Shaft Diameter 3.9 F - 4.2 F*
Guidewire Compatibility 0.018"

*Depending on balloon size

ORDERING INFORMATION

Reference Number
Usable Length
90 cm
Reference Number
Usable Length
130 cm
Reference Number
Usable Length
180 cm
Balloon
Diameter
(mm)
Balloon
Length
(mm)
Recommended
Introducer Sheath
(F)
Nominal
Pressure
(atm)
RBP
(atm)
PCF 040 150 090 PCF 040 150 130 PCF 040 150 180 4.0 150 4 6 14
PCF 040 200 090 PCF 040 200 130 - 4.0 200 4 6 14
PCF 040 250 090 PCF 040 250 130 - 4.0 250 4 6 14
PCF 040 300 090 PCF 040 300 130 - 4.0 300 4 6 14
PCF 050 150 090 PCF 050 150 130 PCF 050 150 180 5.0 150 4 6 14
PCF 050 200 090 PCF 050 200 120 - 5.0 200 4 6 14
PCF 050 250 090 PCF 050 250 130 - 5.0 250 5 6 14
PCF 050 300 090 PCF 050 300 130 - 5.0 300 5 6 14
PCF 060 150 090 PCF 060 120 130 - 6.0 150 5 6 12
PCF 060 200 090 PCF 060 200 130 - 6.0 200 5 6 12
PCF 060 250 090 PCF 060 250 130 - 6.0 250 5 6 12
PCF 060 300 090 PCF 060 300 130 - 6.0 300 5 6 12
PCF 070 150 090 PCF 070 150 130 - 7.0 150 5 6 12
PCF 070 200 090 PCF 070 200 130 - 7.0 200 5 6 12
PCF 070 250 090 PCF 070 250 130 - 7.0 250 5 6 12

 

See product catalogue for complete, detailed product information.

This page may include information about products that may not be available in your region or country. Please consult the approved indications for use. Content on specific Medtronic products is not intended for users in markets that do not have authorisation for use.

*

Information above is supported by Bench test data which is on file at Medtronic, Inc. Test data is not indicative of clinical performance.

4 F/5 F depending on balloon size