QUANTUM HEATER COOLER DISPOSABLES: HEAT EXCHANGERS

Standard Quantum Heat Exchangers

Low flow (Neonatal)

  • Priming volume blood side: 11 ml
  • Priming volume HTF side: 22 ml
  • Exchange surface: 0.028 m2
  • Maximum rate: 1 l/min
  • Blood connections: 3/16’’ barbed profile
Standard Heat Exchanger Low Flow Neonatal

Medium flow (Pediatric)

  • Priming volume blood side: 22 ml
  • Priming volume HTF side: 38 ml
  • Exchange surface: 0.05 m2
  • Maximum rate: 3 l/min
  • Blood connections: 1/4’’ barbed profile
Standard Heat Exchanger Medium Flow Pediatric

High flow (Adult)

  • Priming volume blood side: 55 ml
  • Priming volume HTF side: 70 ml
  • Exchange surface: 0.12 m2
  • Maximum rate: 8 l/min
  • Blood connctions: 3/8’’ barbed profile
Standard Heat Exchanger High Flow Adult

Cardioplegia Quantum Heat Exchangers

Low flow (Pediatric and Neonatal)

  • Priming volume blood side: 22 ml
  • Priming volume HTF side: 22 ml
  • Exchange surface: 0.028 m2
  • Maximum rate: 0.5 l/min
  • Relief valve opening pressure: 500 mmHg
  • Blood connections: 3/16’’ barbed profile
Cardioplegia Heat Exchanger Low Flow Pediatric
cardioplegia-heat-exchanger-low-flow

High flow (Adult)

  • Priming volume blood side: 32 ml
  • Priming volume HTF side: 38 ml
  • Exchange surface: 0.05 m2
  • Maximum rate: 1 l/min
  • Relief valve opening pressure: 500 mmHG
  • Blood connections: 1/ 4’’ barbed profile
Cardioplegia Heat Exchanger High Flow Adult
cardioplegia-heat-exchanger-high-flow

request  more information

CONTACT US

DISCLAIMER    